Macurco CD-6G Operation Manual
If the error mode repeats frequently, check for continuous power and proper voltage. If power is not the problem
and a unit has repeating error conditions, it may need to be returned to Macurco for service, per these User
Instructions.
N ote
: If multiple error codes are occurring simultaneously, the different error codes will be added together into one
4 2 T
4 2 T
combined error code. Example: t08 and T80 occurring simultaneously will display as t88.
If the error mode indicates "Sensor expired" see section
5.2 Sensor Poisons
Unlike traditional catalytic bead and electrochemical sensors, IR sensors use a light source and a receptor to detect a specific
gas. Therefore, the sensor less likely to be affected by inhibitors or poison significantly enhancing safety and sensor life.
5.3 End-of-Life Signal
The CD-6G has a long life, the non-replaceable infrared sensor (IR Sensor). Fifteen (15) years after the CD-6G is installed the
sensor end-of-life signal will be activated indicating that the CD-6G has reached the end of its typical usable life. The end-of-life
signal will cause an error code t200 "Sensor expired". See section
48 hours by pressing the "ENTER/TEST" button or by temporarily dropping power to the unit. The end-of-life signal provides the
user an opportunity to test and/or calibrate the sensor assuring that it is still performing within acceptable parameters though
the sensor is nearing the end of its expected life. The silence function will continue to be available for 29 days after the CD-6G
initiates the initial end-of-life signal. After this 29-day period, the CD-6G can no longer be silenced, and the unit must be
replaced.
Do not disassemble unit or attempt to repair or modify any component of this instrument. This
instrument contains no user serviceable parts, and substitution of components may impair
product performance.
6 Maintenance
The CD-6G is low maintenance. The unit uses a long life NDIR sensor that has a 15-year life expectancy (in normal conditions).
The CD-6G does not perform an automated background calibration and will need to be bump tested and calibrated to verify
functionality and accuracy of the sensor.
N ote:
This applies when used in the typical indoor ambient air with weekly unoccupied periods.
4 2 T
4 2 T
All maintenance and repair of products manufactured by Macurco are to be performed at the appropriate Macurco
manufacturing facility. Macurco does not sanction any third-party repair facilities.
NOTE: CD-6G must be tested at regular intervals in accordance with the requirements of the National Fire Protection
Association (NFPA) 720.
REV – 1.0
5 .3 End-of-life Signal
3 1 T
5 .1.2 "t" Error Codes
3 1 T
WARNING
[34-2900-0510-9 ]
of these User Instructions.
3 1 T
.
The end-of-life signal can be silenced for
3 1 T 3 1 T
3 1 T
22
|
P a g e