Section 7 Troubleshooting
7-6
Inadequate pump pressure
Ensure any user installed in-line valves are in the desired position.
Ensure the circulator's process fluid outlet is connected to the
application's fluid inlet and not the outlet, see Section 3.
Ensure all connections are secure and that the proper sealant/lubricant
for the fitting material is used.
Keep the distance between the circulator and the application as short as
possible.
Ensure tubing is straight and without bends. If diameter reductions are
required, make them at the inlet and outlet of your application, not at
the circulator.
Circulator shuts down
Ensure
button wasn't accidently pressed.
Ensure the circuit protector(s) is in the on ( I ) position.
Check the display for error codes.
Make sure supply voltage is connected and matches the nameplate
rating ±10%.
Restart the circulator.
USB Driver Not Recognized
If your operating system does not automatically recognize the optional
driver log on to:
http://www.ftdichip.com/FTDrivers.htm
for instructions.
Please contact Thermo Fisher Scientific Sales Service and Customer
Support if you need any additional information, see inside cover for
contact instructions.