Guidance and manufacturer's declaration - electromagnetic immunity
The m246 is intended for use in the electromagnetic environment specified below.
The customer or the user of the m246 should assure that it is used in such an environment.
IMMUNITY test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
For all ME EQUIPMENT and ME SYSTEMS
IEC 60601
test level
± 6 kV contact
± 8 kV air
3 A/m
Table 2
Compliance
level
± 6 kV contact
± 8 kV air
3 A/m
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location
in
a
commercial
or
environment.
typical
hospital