For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration - electromagnetic immunity
The m246 is intended for use in the electromagnetic environment specified below.
The customer or the user of the m246 should assure that it is used in such an environment.
IMMUNITY test
Radiated RF
IEC 61000-4-3
Table 3
IEC 60601
Compliance
test level
3 V/m
80 MHz
to 2,5 GHz
Electromagnetic environment -
level
3 V/m
Portable
communications equipment should
be used no closer to any part of the
m246, including cables, than the
recommended separation distance
calculated
applicable to the frequency of the
transmitter.
Recommended separation distance:
d
=
1
d
=
2
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter
manufacturer and d is the
guidance
and
mobile
from
the
P
2 .
80 MHz to 800 MHz
3 .
P
80 MHz to 2,5 GHz
RF
equation