SSCOR S-SCORT newDuet 2314 Manual De Instrucciones De Funcionamiento Y Mantenimiento página 9

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

SSCOR warrants that each new product is free from defects in material and workmanship under nor-
mal use and service for a period of one year from date of purchase.
If returned to SSCOR, we will arrange for repairs or replacement within the terms of the warranty. The
defective product should be decontaminated and returned properly packaged and postage prepaid. Loss
or damage in transit to the factory shall be at the purchaser's risk.
Please call 800.434.5211 - Int. +1 818.504.4054 for return authorization. Loss or damage in return ship-
ment from SSCOR shall be at the purchaser's risk.
The warranty shall not apply to any SSCOR product which has been repaired by anyone other than an
authorized SSCOR representative, or altered in any way so as, in SSCOR's judgment, to affect its safe-
ty or efficacy, nor which has been subject to misuse, negligence, or accident, nor which has had the
serial number altered, effaced or removed.
Neither shall this warranty apply to any SSCOR product which has been connected otherwise than in
accordance with the instructions furnished by SSCOR.
This warranty is in lieu of all other warranties expressed or implied and of all other obligations or lia-
bilities on SSCOR's part, and SSCOR neither assumes, nor authorizes any representative or other per-
sons to assume for it, any other liability in connection with the sale of SSCOR products.
Batteries and disposable collection canisters are excluded from this warranty.
SSCOR will replace any S-SCORT newDUET
specifications for a period of three years from the date of purchase. Verify the condition of the battery
per the weekly checklist on Page 7 in this Operations Manual.
• The unit is designed to suction contaminated fluids, which should be removed from the system
immediately after the procedure. Dispose of the collection canister, patient tube and catheter and
replace. If the procedure produced an excessive quantity of fluids, check the vacuum line (10) for
evidence of moisture. Replace vacuum line if moisture is evident. (Use 3/8" ID tubing.)
• If the vacuum line between the pump and canister is moist, it is possible that fluids have reached the
vacuum pump. Your engineering department will have to open the unit to check the condition of the
pump.
• The capacity of the battery should be checked periodically by running the unit on battery only.
Detach from AC power and run the unit for 15 minutes and look for a blinking LED
(page 7).
• If the Medical Switcher Power Supply (F) appears defective, return the unit to the factory for repair.
Do not attempt to repair the Medical Switcher Power Supply.
• If the Internal Dual Mode Charger (E) appears defective, return the unit to the factory for repair. Do
not attempt to repair the Internal Dual Mode Charger.
• Cleaning: Disconnect AC power source prior to cleaning. Use a mild detergent. Do not use alcohol,
solvents or cleaning solutions. For interior cleaning, use dry, compressed air.
• For technical assistance, call (800) 434.5211, Int. +818.504.4054
Warranty
S-SCORT
newDuet
®
®
Model 2314 Battery Replacement Policy
Model 2314 battery which fails to operate the pump to
®
Maintenance
S-SCORT
®
newDuet
®
9
Model 2314
Model 2314