Cleaning with water
Wash all nebuliser components (except air hose) under warm tap water (max 60 °C) for about 5 minutes adding if necessary
a small quantity of detergent following dosage and use limitations as provided by detergent manufacturer. Rinse thoroughly
making sure that all deposits are washed away and leave to dry.
Disinfecting
All nebuliser components (except air hose) can be disinfected with chemical disinfectants following dosage and use limitations
as provided by disinfectant manufacturer. Disinfectants are usually available at pharmacies.
Sterilizing with steam
All nebuliser components (except air hose and masks) can be heat steam sterilized up to 121 °C (20 min.) or 134 °C (7 min.).
EN554/ISO11134.The sterilization packaging must conform to EN868/ISO11607 and be suitable for steam sterilization.
After sterilization always let all components cool down to ambient temperature before further use.
Do not repeat sterilization cycle when compo-nents are still warm.
MAINTENANCE, CARE, AND SERVICE
Order all spare parts from your dealer or pharmacist.
It is advisable replace nebulizer after some 100 to 120 treatments on single patient or after about 20 sterilization cycles.
Check the filter continually for cleanliness and replace it if dirty, or after a maximum of 70 applications. Spare filters are provi-
ded with the device. To replace the filter 17, lift the filter lid 7 on the bottom of the device, applying leverage with a flathead
screwdriver; remove the filter you wish to replace and put the new one. Then place the filter lid in its housing, making sure it
adheres to the surface.
MALFUNCTIONS AND ACTIONS TO TAKE
The device cannot be switched on
Ensure the power cord 2 is correctly plugged into the socket. Ensure the ON/OFF switch 3 is in the position «I». The device
has automatically switched off due to overheating. Wait until it has cooled down and try again.
The device is on but it does not nebulise
Make sure that the nebulisation nozzle (pisper) 9a in inserted in the vial.
Ensure the air hose 8 is correctly connected at both ends. Ensure the air hose is not squashed, bent, dirty or blocked. If neces-
sary, replace with a new one. Ensure the nebuliser 9 is correctly assembled 14. Make sure that the air intake filter 17 is not
occluded or dirty. In this case, replace it. Make sure that the drug is inside the vial 9.
The device does not work
Thermoprotection intervened for the following reasons: the device was working near heat sources or in environments with
temperature higher than 40°C. Let the device cool off for at least 60 minutes, then turn it on again.
GUARANTEE
This device is covered by a 3 year guarantee from the date of purchase. The guarantee is valid only on presentation of the
guarantee card completed by the dealer (see back) conforming date of purchase or the receipt. The guarantee covers only the
compressor. The replaceable components like nebuliser, masks, mouthpiece, air hose, and filters are not included. Opening or
altering the device invalidates the guarantee.The guarantee does not cover damage caused by improper handling, accidents
or noncompliance with the operating instructions.
0434
TECHNICAL DATA
Piston electrocompressor with thermal cut
Nominal voltage: see data label
Safety fuse: see data label
Dimensions: 196x357x185H mm
Weight: 2.5 Kg approx.
Professional aspiration filter: 50/70 μm
Continuous use compressor
MDD 93/42/EEC - Class of risk IIa
Maximum pressure: 250 kPa
Air flow: 15 Lt/min
Noise level: 55 dBA (EN13544-1)
Neb rate: 0.60 ml/min (tap closed)
Nebulizer capacity: min. 2 ml - max. 16 ml
Operating pressure: from 0 to 110 kPa approx. (suggested range: from 70 to 110 kPa)
Operating flow: from 0 to 6.8 l/min
MMAD: 2.7 μm GSD: 2.98 Aerosol Output: 314 μl Output Rate: 170 μl/min
Operating Temperature/Humidity: +10° C to +40° C/10% to 95% RH
Storage and Transport Temperature/Humidity/Air Pressure:
–25° C to +70° C/
10% to 95% RH/
690 - 1060hPa