Intended Use
This system is a 2 in 1 system designed to measure blood glucose
outside the human body (in vitro diagnostic use) and to measure
blood pressure non-invasively. It is intended for use in the home and in
clinical settings as an aid to monitoring the effectiveness of diabetes
control. It should not be used for the diagnosis of diabetes and hyper-
tension, or for testing neonates.
The alternative site testing in this system can be used only during
steady-state blood glucose conditions.
This monitor has some speaking function but is not intended for use
by the visually impaired.
Test Principle
Your system measures the amount of sugar (glucose) in whole blood.
The glucose testing is based on the measurement of electrical current
generated by the reaction of glucose with the reagent of the strip. The
monitor measures the current, calculates the blood glucose level, and
displays the result. The strength of the current produced by the reac-
tion depends on the amount of glucose in the blood sample.
The blood pressure is measured non-invasively at the arm, based on
the oscillometric method.
This device is NOT able to take measurements in the presence of
common arrhythmia, such as arterial or ventricular premature beats or
arterial fibrillation. It may produce reading errors.