8. Screw in the adapter support (7) hand
tight.
9. The entire instrument incl. membrane
filter is autoclavable at 121 °C (2 bar)
according to DIN EN 285.
For autoclaving remove suction (1) bellow.
Note:
Not every membrane filter is suitable.
Use only genuine manufacturer's recom-
mended accessories. Membrane filters of
a pore size of 3 µm are a standard equip-
ment of the instrument. Up to 5 times,
the membrane filter can withstand steam
sterilization at 121 °C (2 bar absolute, i.e.
1 bar above atmospheric pressure) for
20 minutes. It is the user's responsibility to
ensure effective autoclaving.
28