Moria informs the availability of the One Use-Plus (OUP) handpiece for artificial chamber (#19175) used with One
Use Large-Cut heads (OU LC) (#19184/xxx) for single use or reusable artificial chamber.
1- Indications
OU LC heads for artificial chamber (#19184/xxx) guided for:
• RU (reusable) artificial chamber by the OUP guide ring for ALTK (#19173) mounted on ALTK artificial chamber
(#19161) equipped with its universal cover (#19162)
• SU (single-use) artificial chamber by the OUP LC single-use guide ring (#19187) mounted on single-use
artificial chamber (#19182)
enable to create, with an automated motion, a complete lamellar resection (without hinge) for donor graft
preparation during keratoplasty surgery.
2- Reference & Designation
• Reference: #19175
• Designation: OUP handpiece for artificial chamber
3- Instructions for use
1/ Prepare Evolution 3E control unit (see instruction for use #65090 included in box of OU LC heads, and user
manual Evolution 3E #65060/INTL):
• perform vacuum test on Evolution 3E control unit
• select OUP mode and speed 2
• connect OUP handpiece to Evolution 3E control unit (Fig. 1)
2/ See instruction for use #65090 for preparation of artificial chamber, setting of the graft and preparation
of the head.
3/ OUP handpiece setting:
• screw the head onto the handpiece (Fig. 2)
• inspect the blade under microscope and start the motor to check proper blade oscillation
• set OUP handpiece on OUP guide ring #19173 or #19187 depending on artificial chamber used (Fig. 3)
4/ See instruction for use #65090 for cutting of the graft and end of procedure.
4- Maintenance
For the maintenance of the OUP handpiece for artificial chamber (#19175), please refer to user manuals #65039
or #65040 (§9.3).
As for the cleaning-disinfection-sterilization-storage of OUP handpiece: see chart on following page.
Addendum for user manual #65009
& Instruction for use #65090

Publicidad

loading

Resumen de contenidos para MORIA 19175

  • Página 1 Addendum for user manual #65009 & Instruction for use #65090 Moria informs the availability of the One Use-Plus (OUP) handpiece for artificial chamber (#19175) used with One Use Large-Cut heads (OU LC) (#19184/xxx) for single use or reusable artificial chamber.
  • Página 2: Important Warnings

    Related Steps What to do IMPORTANT WARNINGS picture Cleaning • The handpiece must be carefully Fig.4 cleaned after each surgical procedure. • To clean the screw core of the motor and metallic parts, use the disposable • Do not use abrasives or scrapers to brush (#19149) moistened with cleaning clean any elements of the system.
  • Página 3 4/ Se référer à la notice #65090 pour la découpe du greffon et la fin de la procédure. 4- Maintenance Concernant la maintenance de la pièce à main OUP pour chambre artificielle (#19175) : se référer aux manuels utilisateurs #65039 ou #65040 (§9.3).
  • Página 4: Avertissements

    Étapes Que faire Figure AVERTISSEMENTS Nettoyage • La pièce à main doit être Fig.4 soigneusement nettoyée après chaque • Pour nettoyer le trou fileté du moteur procédure chirurgicale. ainsi que les parties métalliques, utiliser la • Ne pas utiliser d’abrasifs ni de grattoirs brosse jetable (#19149) humidifiée avec pour nettoyer les divers éléments.
  • Página 5: Referencia Y Designación

    4/ Consulte las instrucciones de uso #65090 para cortar el injerto y finalizar el procedimiento. 4- Maintenance Para el mantenimiento de la pieza manual OUP para cámara artificial (#19175), por favor consulte los manuales del usuario #65039 o #65040 (§9.3).
  • Página 6: Avisos Importantes

    Pasos Qué hacer Imagen AVISOS IMPORTANTES Limpieza • La pieza manual debe limpiarse Fig.4 cuidadosamente después de cada • Para limpiar el tornillo del motor y procedimiento quirúrgico. las partes metálicas, utilice el cepillo desechable (#19149) humedecido con • No utilice abrasivos o espátulas para soluciones de limpieza.
  • Página 7 Fig.1 Fig.2 Fig.3 Fig.4      ...
  • Página 8 To obtain more information MORIA SA MORIA Inc MORIA JAPAN KK Moria Shanghai Office 15, rue Georges Besse 1050 Cross Keys Drive Inami Bldg. 6F RM4, 10 FL, Kai Li Building 92160 Antony Doylestown, PA 18902 3-24-2 Hongo Bunkyo-ku N°432 Huai Hai West Road...