Chapter 5. Operation; General Use - B.A. International Optima OS600L Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 17

Chapter 5. Operation

5.1 General use

1) Turn on the Power switch of implant engine controller
2) Programs are selected in turn by pressing Foot switch or P button of implant engine controller.
3) Check the displayed torque, rotation speed, irrigation flow rate, gear ratio and direction of rotation.
4) It starts rotating upon pressing Foot switch. It rotates at low speed with light press on Foot switch and rotates at full speed with hard press.
When the irrigation flow rate is preset, the pump starts rotating as well.
5) When load reaches the maximum value of preset torque, motor stops rotating.
6) It stops rotating upon releasing Foot switch.
7) Turn off the implant engine controller
① Connect the engine to the power cable.
② Turn the engine on at the power switch.
① When not in use, turn off the engine at the power switch
② Disconnect the engine from the power cable