Apply a flame to the nozzle elbow to melt the wax accumulated inside. Tilt the burner to allow the
wax to flow out freely through the flame nozzle.
After removing the wax, apply a ¼ turn to the flame nozzle to return to the initial position, and
connect the burner to the gas mains as described above, in reverse order.
PRECAUTIONS
♦ Before connecting the burner, ensure that the gas supply is correct (butane, natural gas).
♦ Only connect the unit to a gas supply which has undergone official inspections performed by licensed
technicians.
♦ Make sure that the rubber hose and connection clasp of the burner to the gas mains are solid and
secure.
♦ Never allow children or unqualified staff to use the burner.
♦ Make sure the flame has a bright blue base and a strong yellow tip. Reddish colors, and black smoke
in particular, indicate poor combustion. Always avoid this situation using the airflow valve.
♦ After using the burner, always close the shutoff cock and gas mains stopcock.
♦ Always position the burner in your laboratory well away from inflammable products.
♦ Never leave the burner operating unattended.
♦ If you smell gas, close the mains stopcock immediately and inform a licensed technician.