Descargar Imprimir esta página

KCI Prevena 125 Guía Para El Paciente página 2

Sistema de tratamiento para incisiones

Publicidad

Idiomas disponibles

Idiomas disponibles

Therapy On/Off
!
Green LED Indicates
Therapy is On
Leak Alert
!
x1
Check all connectors
for leaks.
Low Battery Alert
!
x1
A slow beep/flashing LED = Change batteries within six hours.
A rapid beep/rapidly flashing LED = Change batteries Immediately.
Maximum Capacity Alert
!
x2
Perform visual inspection of canister.
System Error Alert
!
!
x1
x1
Hold 5 seconds to TURN OFF.
Dressing Pressure Indicator
UP Position
Indicators and Alerts
TURN ON
TURN OFF
Hold 2 seconds.
Hold 5 seconds.
!
!
Ensure canister is securely locked
Press firmly around
and side tabs are flush with unit.
dressing edge.
Full or near full?
!
Hold 2 seconds to TURN ON.
Check for leaks
See Leak Alert
Check for kinked
section above.
or pinched tubing.
AUDIO PAUSE
Push once to mute Alerts
Device Life-cycle
for 60 minutes.
Expired
!
Life Cycle 192 hours (8 Days).
Eliminated?
Use Prevena™
Patch Strips to seal leak.
!
3 "AA" Batter
Observe (+) and (-)
for orientation.
Hold 5 seconds
to TURN OFF.
NO
Check for kinked or pinched tubing.
YES
Turn therapy unit off. Call Treating Physician immediately.
See WARNINGS Section on opposite page.
Therapy Running?
Indicator Down?
!
Continue Therapy.
YES
Leak Alert
NO
ies
(Lithium Batteries
Recommended).
Continue Therapy.
YES
NO
Call KCI at 1-800-275-4524.
YES
Continue Therapy.
Call KCI at
NO
1-800-275-4524.
Call KCI at
1-800-275-4524.
Call KCI at
1-800-275-4524.
!
Hold 2 seconds