Principles of Operation

+
Each Cimarec
unit utilizes the latest microprocessor
technology to deliver a reliable, controlled, ceramic top
stirring hot plate.
+
Your Cimarec
stirring hot plate has an electronic closed-
loop feedback control which will accurately maintain
temperature setpoints from 30°C through 300°C, 400°C
or 540°C, depending on the model. Top plate surface
temperatures are tested using the latest infrared
temperature measurement. The measurements are
made with an infrared camera mounted approximately
26" above the top surface of the stirring hot plate. If the
temperature measurement of the ceramic top is made
with measurement devices other than infrared, the error
of the measuring technique may be greater than the
error of the unit.
The electronic stirring speed control will maintain the
speed setpoint when the unit is loaded at ±5.0%. The
+
motor in the Cimarec
stirring hot plates produces maxi-
mum stirring torque under normal laboratory load con-
ditions, and is combined with a powerful magnet to
provide exceptional magnetic coupling with a stir bar.
When stirring is turned off, the unique braking feature on
+
the Cimarec
models immediately brings the fluid and
stir bar down to a safe speed for quick flask removal.

General Cleaning Instructions

Wipe exterior surfaces with lightly dampened cloth
containing mild soap solution.