Troubleshooting

Error Codes

Errors E01 through E04 are heating errors. Error Handler will lock out heating functions if heating error is detected.
Stirring functionality is unaffected. Error E05 is stirring error. Error Handler will lock out stirring functions if stirring error
is detected. Heating functionality is unaffected. If the condition that caused the error is no longer present, the unit will
clear Errors E01-E05 automatically.
Displayed Message Intended to Detect
E01
E02
E03
E04
E05
28
Internal thermocouple out
of range.
Excessive top heat-up
time.
External Probe left out of
solution.
OTP thermocouple out of
range.
Motor system failure.
Cause
Internal thermocouple not
connected.
Thermocouple open.
Thermocouple connected
backwards (reversed polarity).
Internal thermocouple short circuit. Remove short.
Failure in Internal thermocouple.
Failure in Element.
Failure in optocoupler/triac circuit. Replace Control Board.
External Probe left out of solution. Place external probe into
External probe connected
backwards.
OTP Circuit failure.
Locked rotor condition.
Failure of motor.
Failure of motor circuit.
Failure of motor sensor.
Solution
Ensure proper connection
and polarity of thermocouple.
Replace thermocouple
(attached to element).
Ensure proper connection
and polarity of thermocouple.
Replace thermocouple
(attached to element).
Replace Element.
solution.
Correct orientation of external
probe.
Replace Control Board.
Free locked rotor.
Replace motor.
Replace Control Board.
Replace motor sensor.