5027 Sampler
be correct. There will be a space between racks 1
and 4.
8.3
Operation
The sampler can be used together with various
analysis systems. This section describes how the
various systems are used together with the
sampler.
Warning
The responsible body shall be made
aware that, if the equipment is used in
a manner not specified by the
manufacturer, the protection
provided by the equipment may be
impaired.
Modifications, alterations, rebuilding
or use of safety parts not authorized
by FOSS Analytical AB violates the
warranty. FOSS Analytical AB has no
responsibility for damages, material
or personal, occuring as a result of
such actions.
8.3.1
Connecting Analysis Sys-
tems
The connector cable with which you will connect
the sampler to the analyzer is usually delivered
together with the analysis system.
8.3.2
Identification of
Operating Controls
1
2
4
3
Fig. 8:1
Operating controls
1
Mains switch
2
Mains select
3
Mode switch
4
Reset button
5
Rocker switch (Functions only in
mode 4 and 5)
8.3.3
Setting the Analysis
Configuration
On the rear of the sampler there is a switch
marked "Mode". The "Mode" switch is a
hexadecimal switch with which you can select
various functions for the sampler, such as
different numbers of sample cups or set it up to
different analysis systems, see Fig. 8:2.
The "F" switch is a neutral position to be used
when the sampler is to be transported or inspected
by service personnel. You can change the position
of the switch using a small screwdriver. Use the
table below to select the appropriate mode of
operation:
8:2
Mode
Configuration
0
Scaba 5600
3
MilkoScan FT 120
4
FIAstar 5010 and
Aquatec 5400
5
FIAstar 5010 and
Aquatec 5400
9
WineScan and
FIAstar 5000
C
FIAstar 5012, Scaba
5610 and 5611
D
FIAstar 5000,
FIAstar 5012 and
Wine-scan
F
Test and adjustment
Note:
If you change the position of the
switch, the sampler must be reset.
5
5027023A
Fig. 8:2
The
8.3.4
Starting the Sampler
Connect the mains cable to the wall socket and
start the sampler by pressing the mains switch on
the rear. The green LED at the front lights up
when the power is on. The sample probe will
automatically move to the wash bottle if it
accidentally moved out of this position.
8.3.5
Operation
Use the table in section 8.3.3 Setting the Analysis
Configuration, to set up the mode switch in the
desired position. Remember to restart the sampler
for a change to take effect. Usually the PC
software of the different analysis systems include
a functional or diagnostic test of the sampler, if
you wish to check it further.
8.4
Instructions for cleaning
Clean unit and sample racks using a mild
detergent.
User Manual 5000 3694 / Rev. 3
Trays Logic
32
TTL
56
RS232
32
TTL
120
TTL
64
RS232
32
RS232
120
RS232
TTL
5027028a
"
"
MODE
switch