For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF
communications equipment and the 04-825-001.
The 04-825-001 is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the 04-825-001
can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the
04-825-001 as recommended below, according to the maximum output power of the
communications equipment.
Separation distance according to frequency of transmitter.
Rated maximum
150 kHz to 80 MHz
output power of
transmitter W
d=1.2 P
0,01
0.12
0,1
0.38
1
1.2
10
3.8
100
12
04-825-001 IM_13Feb2015.indd 30-31
Table 4
80 MHz to 800
800 MHz to 2.5
MHz
GHz
d=1.2 P
d=1.2 P
0.12
0.23
0.38
0.73
1.2
2.3
3.8
7.3
12
23
30
Table 4 - Continued
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be determined using the equation applicable
to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people
and animals.