Guidance 3
... ...
... T he i700 is intended for use in the electromagnetic environment
specified below. The customer or the user of the i700 should
assure that it is used in such an environment.
IEC
Immunity
60601
Compliance
Test
test
level
3 Vrms
150 kHz
– 80 MHz
Outside
Conducted
ISM
RF IEC
Bands
6
c)
61000-4-6
Vrms 150
kHz – 80
MHz In ISM
Bands
c)
20
User Manual
Recommended
Electromagnetic
separation
environment
level
distance(d)
‐ guidance
Portable and mobile
RF communications
equipment, including
cables, should be used
no closer to any part
of the i700 than the
3Vrms
d = 1.2√P
recommended separation
distance as calculated
using the equation
below, according to
the frequency of the
transmitter.
Radiated
3 V/m 80
6 V/m
RF
MHz
IEC 61000-
to 2.7 GHz
4-3
IEC
Where P is the maximum
60601‐1‐
output power rating
2:2007 d = 1.2
of the transmitter in
√P 80 MHz to
watts (W) according
to the transmitter
800 MHz d =
2.3√P 80 MHz
manufacturer and d is the
to 2.5 GHz
recommended separation
distance in meters (m).
IEC 60601‐1
Field strengths from
‐2:2014 d=2.0
fixed RF transmitters,
√P 80 MHz to
as determined by an
2.7 GHz
electromagnetic site
survey
should be less
a)
than the compliance level
in each frequency range
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
b)