Programming Chart - DJO CEFAR REHAB X2 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Programming chart

24
Level 1
Level 3
Conventional
(C)
Frequency
Preset: 80 Hz
Range: 1–120 Hz
Pulse duration
Preset: 180 µs
Range: 50–400 µs
Time
Preset: – – min.
Range: 1-99, – – min.
Programming
CEFAR REHAB X2
P28, P29, P30
CONTINUOUS
(C)
See the programming chart on next
Burst
Pulse width/duration
(B)
Frequency
Preset: 2 Hz
Range: 1–5 Hz
Time
Preset: – – min.
Range: 1-99, – – min.
Range: 1-99, – – min.
INTERMITTENT
(I)
page for intermittent stimulation
modulation
(PWM)
Frequency
Pulse duration
Preset: 80 Hz
Preset: 180 µs
Range: 2–120 Hz
Range: 50–400 µs
Pulse duration 1
Preset: 70 µs
Range: 2–120 Hz
Range: 50–400 µs
Pulse duration 2
Preset: 180 µs
Range: 50–400 µs
Range: 2–120 Hz
Modulation time
Modulation time
Preset: 3 s
Range: 1–10 s
Range: 5–15 s
Time
Preset: – – min.
Preset: – – min.
Range: 1-99, – – min.
Frequency
modulation
(FM)
Frequency 1
Preset: 5 Hz
Frequency 2
Preset: 15 Hz
Preset: 5 s
Time

Hide quick links:

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

2946

Tabla de contenido