BP 603 GB_neu
04.01.2005
11 an error is displayed
Error codes:
Err 0.:
No pulse or the pulse measured is too low.
Err 1.:
The sleeve is not airtight or the pressure build-up is too slight.
Err 2.:
The systolic value cannot be obtained.
Err 3.:
The sleeve pressure is too high.
Err 4.:
The diastolic value cannot be recorded or is outside the measuring range.
Err 5.:
The pressure reduction occurs too quickly.
Err 6.:
The pressure reduction occurs too slowly.
Err 7.:
The systolic value is outside the measuring range.
Err 8.:
The sleeve pressure is outside the measuring range.
Err 9.:
The sleeve pressure is insufficient, although highest point of the pressure build-up has
already been reached.
22
All manuals and user guides at all-guides.com
10:45 Uhr
Seite 22