1. To view the calibration data that is stored in a probe, push the
and select View Probe Data. The current calibration and calibration
history for the probe can be viewed.
Option
Description
View Current
The current calibration information shows the calibration
Calibration
details for the most recent calibration. If the probe has not
been calibrated by the user, the factory calibration data is
shown.
View Calibration
The calibration history shows a list of the times when the
History
probe was calibrated. Select a date and time to view a
summary of the calibration data.
Print stored data
The meter must connect to AC power to start the USB connection. Make
sure that the connection to AC power is made before the meter is powered
on.
All data can be sent to a printer. The PD-24 Citizen Printer is compatible
with the HQd meters and is FCC Part 15B, Class B compliant with the HQ
meters. Other printers may not be compatible. Compatible printers should
support a minimum of 72 columns of data, be capable of printing up to 500
continuous data-stream events in 1, 2 and 3 lines of text and fully support
code page 437 and code page 850.
1. Turn off the meter. Make sure that the meter is connected to AC power.
Refer to
Connect to AC power
2. Connect the printer to the meter with a USB cable type A. Refer to
Figure
9.
3. Turn on the meter.
4. Push the
key.
5. Select Send Data Log. Wait for the display to show "Transfer
Complete" and wait for the printer to stop printing. Disconnect the
printer.
on page 5.
Figure 9 Connection to the printer
key
1 USB cable
2 Citizen Printer, FCC Part 15B, Class
B compliant
Change the report options
Printed reports for sample data can contain 1, 2 or 3 lines of information.
For additional information, refer to the CD.
1. Push the
key. Select Report Options.
2. Select Report Type and select one of the options.
Option
Description
Basic report
One line of data.
3 AC-DC power supply for printer