Fitting The Device - Össur PROPRIO FOOT W/UNITY Instrucciones Para El Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 16
Precautions
• U.S. federal law requires that the device is sold by a prosthetist and by the order of a physician. The practitioner
must provide each user with training for appropriate use and proper application of the device.
• The device is to be used exclusively for unilateral or bilateral major lower-limb amputations.
• Use a Unity
Socket Valve Kit to connect the device to the socket.
®
• Ensure the socket is a close fit to prevent blistering and/or wounds.
• Ensure the tube is not folded, bent, or closed off in any way (Figure 3).
• Ensure the tube is secured firmly away from any sharp objects or components.
N.B. Practitioners must ensure their clients are aware of, and understand the following:
• The tube must always be well secured.
• The device must never be used if a change in its operation is observed. In such cases, the user should consult
their practitioner immediately.
Weatherproof
A weatherproof device allows use in a wet and/or humid environment, but does not allow submersion. Fresh water
splashing against the enclosure from any direction shall have no harmful effect. Dry thoroughly after contact with
fresh water. Fresh water: Includes tap water. Excludes salt and chlorinated water.
For a weatherproof system, weatherproof components must be selected for use with the device.
STRUCTURAL SPECIFICATIONS
Net weight: Approximately 80g (including tube).
Frame construction: Carbon fiber and aluminum.
Clearance and dimensions: Refer to the PROPRIO FOOT

FITTING THE DEVICE

Warning: Ensure the socket is securely fit before using the device; never use with a loose fitting or leaking socket.
N.B. To connect the system to a socket, a Unity
the sole responsibility of the provider.
1. Fit the user with a prosthetic foot containing the device.
2. Cut the tube to correct length, and connect to the device and socket (using the Unity
3. Use the supplied tube fixture to secure the tube against the pylon. The tube should be located medially rather
than laterally to avoid damage during ambulation.
4. Once the device has been connected, ensure the pump is creating a vacuum during gait and not leaking air.
Retrof itting the Device
Use an extended ball-end hex size 4mm torque screwdriver to 3Nm.
THE TUBE
The tube is the most sensitive part of the device, and special attention should be paid to it for potential functional
issues that might arise.
Precaution: Ensure that the device tube is not compressed or folded by the application of cosmetic finishing.
Ensure the tube is NOT:
• Folded (Figure 3).
• Pressed up against sharp objects.
• Punctured.
• Dangling during gait.
• Located laterally, increasing likelihood of damage.
FINISHING / LINER OPTIONS
Össur recommends that the device is used with an Iceross Seal-In liner.
Information For Use.
®
Socket Valve Kit is recommended. The use of other applications is
®
Socket Valve Kit).
®
5

Publicidad

Tabla de contenido
loading

Tabla de contenido