4.2 Filter change
In normal operating conditions, the compressor filter must be
In normal operating conditions, ...
replaced after every 200 operating hours, but at least once a
year.
Also please check the filter at regular intervals (after every
10 inhalations). If it is soiled (grey or brown colour) or blocked,
replace it. The filter must also be replaced with a new one if it has
become damp.
The filter cannot be cleaned and then reused!
Only use ...
NOTICE
Only use air filters that are intended by PARI for use with your
compressor, otherwise your compressor might be damaged.
To replace the filter, proceed as follows:
• Unscrew the filter holder from your
compressor using a suitable coin.
• Pull the old filter out of the filter holder and fit the
new filter in its place.
• Reinsert the filter holder together with the new
filter in your compressor and screw it tight with
the coin.
- 40 -