Components
Overview (fig. I)
1 Handle
2 LCD
3 Infrared sensor
4 Actuator (measurement button)
5 Battery cover
Indicator
6 Temperature values for MAX
7 Symbol for MAX
8 Currently measured value
9 Display of measured value
10 Data hold
11 Symbol laser on/off
12 Symbol lock
13 High/low alarm symbols (HAL; LOW)
14 Symbols for °C / °F
15 Low battery symbol
16 Symbol and value for emissivity
Keys
17 Down key (lighting (on/off))
18 Mode key (menu loop)
19 Up key (laser on/off)

Function overview

• Precise non-contact measurements
• The excellent optical system allows measure-
ments even from long distances
• Wide temperature measurement range
• Double laser pointer
• MAX-HOLD-LOCK function
• Adjustable emissivity
• Optical and acoustical alarm if the tempera-
ture falls below the freely adjustable values or
if it exceeds them
• Backlight
• Possibility to select °C/°F

Technical specifications

Item
Measuring range
Optical resolution D:S
Display resolution
Precision
(ambient temperature from 23°C - 25°C)
-50 °C ~ 20 °C
20 °C ~ 300 °C
300 °C ~ 650 °C
Reproducibility
-50 °C ~ 20 °C
20 °C ~ 650 °C
Response time
Spectral responsitivity
Operating temperature
Storage temperature
Laser
Emissivity
Power supply
Relative humidity
10 % ~ 90 % RH (operation)
< 80 % RH (storage)
0715 53 110
-50 °C - 650 °C
(-58 °F - 1202 °F)
12:1
0.1 °C
± 2.5 °C
± 1.0 % ± 1.0 °C
± 1.5 %
± 1.3 °C
± 0.5 °C
150 ms
8 ~ 14 μm
0 - 50 °C
-10 - 60 °C
Class II
adjustable from 0.10
to 1.0
1 x 9 V battery
11