The maximum allowed load (evaporation flask and contents) is 3 kg.
Filling medium by vacuum: prior to filling the evaporation flask, a vacuum controller is used to regulate the
glass apparatus to the target pressure.
- Now fill the evaporation flask using the backfeed line.
- Due to the vacuum present, the solvent is suctioned into the evaporation flask. This enables you to keep
solvent loss due to suctioning to a minimum.
Setting up the heating bath:
Also refer to the operating instructions for the HB 10 heating bath!
- Move the lift to the bottom position and check the position of the heating bath in relation to the evaporation
flask. When using larger evaporation flasks (2 or 3 liters) or depending on the angle of the rotation drive, you
can move the heating bath to the right.
- Fill the heating bath with the tempering medium until the evaporation flask is surrounded by tempering
medium to 2/3 of its volume.
- Switch on the rotation drive and slowly increase the speed.
Note: Avoid creating waves.
- Switch on the heating bath using the main on/off switch.
Note: Avoid stress on the glass due to different evaporation flask and heating bath temperatures when
lowering the evaporation flask into the heating bath!
Lift position key (4&6)
To adjust the lift position, grasp the handle and
press the left or right lock button (4).
The hand must rest on the sensor (6).
Note: Hand detection sensor (6), only to be
touched by hand, do not press!
The lift lock is unlocked and you can move the
lift up and down.
Release the lock button or remove the hand
from the lift, the lift is locked in the adjusted
position.
When locking and unlocking, a clear "click"
sound is heard.