Charging The Rechargeable Battery - Dräger UCF FireVista Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 25
Error
Cause
LED in the charging cra-
Camera is not seated
dle indicates an error
correctly in the charging
during the charging pro-
cradle.
cess or nothing at all.
Contacts on the camera
and/or the charging cra-
dle are dirty.
Contacts on the camera
and/or the charging cra-
dle are damaged.
The retainer on the
charging cradle is not
closed.
No power supplied to the
charging cradle.
6
Maintenance
The following components of the thermal imaging camera in particular can be
replaced by DrägerService:
Rechargeable battery
Window of the IR camera
Display
Instructions for use
|
Dräger UCF FireVista
6.1
Remedy
The display and window of the IR camera consist of scratch-resistant material.
Check the position of the
Nevertheless, take care when cleaning these optical surfaces.
hand straps or the carry-
ing strap and correct if
NOTICE
necessary.
The camera housing could be damaged if solvents or thinners are used for
Remove foreign objects
cleaning.
from under the camera
► Do not use any solvents or thinners.
or in the charging cradle.
1. Wipe the thermal imaging camera with a damp, lint-free cloth. If necessary,
Place charging cradle on
clean with warm, soapy water.
level surface.
2. Dry the thermal imaging camera with a soft, lint-free cloth, and leave to dry
completely.
Clean the charging pins
on the camera and the
charging cradle.
6.2
If the charging pins are
defective, send the cam-
NOTICE
era and the charging
If the charging cradle comes into contact with water, there is a risk of a short
cradle to Dräger.
circuit.
Close the retainer.
► Make sure that the thermal imaging camera and the charging cradle are dry.
Charge the thermal imaging camera in ambient temperatures of 5 °C to
40 °C. The charger stops charging and indicates an error to prevent damage to
Check the functioning of
the rechargeable battery when temperatures are outside this range. The
the power supply unit
charger continues charging as soon as the camera temperature reaches the
and the connecting
permissible range again.
cable.
The thermal imaging camera can be returned to DrägerService when the
rechargeable battery can no longer be charged. DrägerService can change the
rechargeable battery.
1. Place the thermal imaging camera in the charging cradle. Close the retainer
on the charging cradle.
 The LED flashes green when the rechargeable battery in the thermal
 The LED lights green when the rechargeable battery in the thermal
 The LED rapidly flashes red if an error has occurred.
Cleaning the thermal imaging camera

Charging the rechargeable battery

imaging camera is being charged.
imaging camera is fully charged.
|
en
Maintenance
15

Publicidad

Tabla de contenido
loading

Tabla de contenido