Operating Instructions
Specifications
Internal transmission fluid
Power supply L+
Signal output and
maximum ohmic load R
A
Setting time (Analogue signal)
Current consumption
Total current consumption
Switch points
Type
Number
Function
Contact rating
Switching current
Response time
Accuracy
Insulationvoltage
Display
Design
Range
Accuracy
Update
Accuracy
Non-linearity
28
© SICK AG • Germany • All rights reserved • Subject to change without notice
7. Starting, operation
Model PBS
Synthetic Oil
(only for pressure ranges < 0 ... 10 bar and ≤ 0 ... 25 bar abs)
L+ in V DC
15 ... 35
R
in Ohm
4 ... 20 mA, 3-wire
A
0 ...10 V, 3-wire
Adjustment zero point offset, max. 3 % of span
ms
3
mA
max. 100
mA
max. 600 (max. 500 with IO-Link) incl. switching current
Individually adjustable via external control keys
Transistor switching output PNP or NPN, with IO-Link: C/Q1: PNP
1 or 2
normally open / normally closed; windows- and hysteresis function freely
adjustable
V DC
Supply voltage L+ - 1 V
mA
Q1: 250 (with IO-Link: C/Q1: 100), Q2: 250
≤ 10 (20 ms at IO-Link)
ms
% of span
≤ 0.5 (setting accuracy)
V DC
500
14-Segment-LED, blue 4-digits, height 9 mm
Electronic 180° rotatable
% of span
≤ 1.0 ± 1 Digit
ms
1000, 500, 200, 100 (adjustable)
% of span
≤ 1.0 *
)
*
Including non-linearity, hysteresis, zero point and full scale error (corresponds to
)
error of measurement per IEC 61298-2)
≤ ± 0.5
% of span
≤ 0,5 k
R
A
R
> 10 k
A
(BFSL) according to IEC 61298-2
8013154/UP90/2015-03-30
PBS