Charging; Connecting; Noting The Serial Number; Troubleshooting - signia Styletto X Guía Del Especialista

Ocultar thumbs Ver también para Styletto X:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Charging

Charge the hearing instruments for at least 15 minutes 
before a first fitting session.
Fully charge the hearing instruments before delivery 
to the client. A full charging cycle takes approximately
4 hours.
Follow the operating conditions in both the hearing
instrument's and the charger's user guide.

Connecting

Use Noahlink Wireless to fit the hearing instruments 
wirelessly.
Noting the serial number
The serial number is printed on the
housing.
6

Troubleshooting

Before sending a hearing instrument to our service 
address, perform a visual check to see if the receiver is
blocked by cerumen. If this is the case, exchange the
receiver wax guard.
Exchanging the receiver wax guard
Remove the ear piece.
X
Screw the threaded end of the
X
wax guard changing tool into
the filter of the external receiver.
Remove the used filter.
X
Flip the tool over and push the
X
new wax guard into place.
Release the filter by tilting the 
X
changing tool to the side. Then
dispose of the changing tool.
7

Publicidad

Capítulos

Tabla de contenido

Solución de problemas

loading

Tabla de contenido