This message displays in various cases – please find below a list of the main cases:
Reasons for LO
message display
Temperature reading hampered by hair,
perspiration...
Temperature hampered by an air flux.
Temperature readings too close together, the
ThermoFlash® LX-260T EVOLUTION did not
have the chance to boot itself.
The measuring distance is too far.
24
Visiomed - LX-260TE - 032017
Advice
Make sure that there is no obstruction prior to
taking a temperature.
Make sure there is no air flux as this could in-
terfere with the infrared system.
Respect the pause of 15 seconds minimum bet-
ween two readings – 1 minute pause is advised.
Please respect the measuring distance (bet-
ween 2 and 5 cm).