Twistorr 305-Ic Mounting - Agilent Technologies TwisTorr 305-IC Manual De Instrucciones

Tabla de contenido

Publicidad

7
Instructions for Use

TwisTorr 305-IC Mounting

TwisTorr 305-IC Mounting

WARNING!
Failure to comply with these installation instructions could result in the pump
detaching from the system in the event of a rotor failure, which could cause
property damage or serious injury or death.
The TwisTorr 305-IC can be installed in any position. Fix the TwisTorr 305-IC in
a stable position, mounting the inlet flange of the turbopump to the system
counter-flange, with a connection capable of withstanding a torque of 1000 Nm
around its axis.
The turbopump with ConFlat and ISO K inlet flange must be fixed to the vacuum
chamber by means of the appropriate Agilent hardware.
For further details see the "Installation of ConFlat and ISO K" section.
The TwisTorr 305-IC cannot be fixed by means of its base
CAUTION!
212/302
TwisTorr 305-IC User Manual / 87-901-058-01 (A.00)

Publicidad

Tabla de contenido
loading

Tabla de contenido