Technical Information

Vent Accessories

Vent Accessories

The vent valve and vent device enable an automatic vent operation.
TwisTorr 305-IC compatible Vent Valve mod. 9699834 – 9699834M006
Figure 38
This vent valve waits before opening a minimum time of about 5 sec. This time can be
increased up to about 220 min. by means of a setting of the controller software
(optional).
258/302
TwisTorr 305-IC User Manual / 87-901-058-01 (A.00)