Specifications
AC Version
Power
100 to 230 V AC, + 15%, 50/60 Hz, 12 VA
(5W) max.
Fuse
Slow-Blow, 0.25 A, 250 V AC
Output
repeatability: 0.25% of full range
resolution: 3 mm (0.1")
relay:
2 Form C (SPDT) contacts,
rated 5A at 250 V AC,
non-inductive
Environmental
• location:
• altitude:
• ambient temperature:
• relative humidity:
• installation category:
• pollution degree:
Process Pressure
• 0.5 bar (7.25 psi) max.
Switching Range
• liquids:
• solids:
Memory
• non-volatile EEPROM
Programming
• 2 keys
Temperature Compensation
• built-in to compensate over the operating range
Display
• LCD
• three 9 mm (0.35") digits for reading of distance between sensor face and material,
multi-segment graphic for operation status
Electronics/Enclosure
• termination:
• material:
• OR
• epoxy coated aluminum with gasket
• ingress protection:
• cable inlet:
7ML19981XB81
DC Version
Power
Output
indoor/outdoor
2000 m max
- 40 to 60 °C (- 40 to 140 °F)
* - 20 °C (-5 °F) if metal mounting
suitable for outdoor (Type 6/NEMA 6/IP67 enclosure)
II
4
0.25 to 5 m (0.8 to16.4 ft)
0.25 to 3 m (0.8 to 9.8 ft)
terminal block, 2.5 mm
stranded, maximum
plastic
Type 6/NEMA 6/IP67
2 x ½" NPT or 2 x PG 13.5
Pointek ULS200 – QUICK START MANUAL
18 to 30 V DC, 3 W
repeatability: 0.25 % of full range
resolution: 3 mm (0.1")
relay:
2 Form C (SPDT)
contacts, rated 5 A
at 48 V DC
OR
transistor: 2 transistor
switches, rated
100 mA maximum at
48 V DC
2
(14 AWG) solid 1.5 mm
1
2
(16 AWG)
Page EN-3

Publicidad

Tabla de contenido
loading

Tabla de contenido