The user is encouraged to try to correct the interference by one or
more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and the receiver.
• Connect the equipment into an outlet on a circuit different from
that to which the receiver is connected.
Consult the dealer or an experienced radio/TV technician for help.
11. Technical Specifications
Model-No.
BM 35
Method of
oscillometric, non-invasive blood pressure
Measurement
measurement on the upper arm
Measurement
cuff pressure range: 0 – 300 mmHg
Range
blood pressure range: 40 – 280 mmHg
heart rate 40 –199 beats/minute
Accuracy of the
systolic ± 3 mmHg / diastolic ± 3 mmHg /
reading
heart rate ± 5% of the displayed value
Measuring
max. permissible standard deviation according
uncertainty
to clinical testing:
systolic: 8 mmHg /diastolic: 8 mmHg
Sets of Memory
2 x 60 memory places (120 memory total)
Unit Dimension
L 5.3" x W 4.1" x H 2.1"
(L 135 mm x W 105 mm x H 53 mm)
Unit Weight
approx. 0.48 lb (217 g)
Cuff Size
upper arm circumference approx. 9" – 13"
(23 – 33 cm)
Note: large cuff (13" to 17" / 33 to 43 cm) is
available at customer service.
Operating
+ 50 °F to 104 °F (+ 10 °C to + 40 °C),
Temperature
15% to 90% RH
Storage- and trans-
- 4 °F to 158 °F (- 20 °C to + 70 °C), ≤ 90% RH
port temperature
Power Supply
4 x 1.5 V AAA-Batteries (alkaline type LR03)
Battery Life
for approx. 250 measurements, depending on
blood pressure level or inflation pressure
Accessories
storage bag, operating instructions,
4 AAA batteries
BF Classification
Key to symbols
application part Type BF
Discard the used product to the recyc-
ling collection point according to local
regulations.
Auto-shut-off
1 minute after last button operation
These specifications are subject to change without notice for pur-
pose of improvement.
• This unit is in line with the International Standard
IEC 60601-1-2 and is subject to particular pre-
cautions with regard to electromagnetic compat-
ibility (EMC). Please note that portable and mobile
17
– Internally powered equipment
– BF type applied part
– IPXO
– Not suitable for use in presence of flam-
mable anesthetic mixture with air or with
Oxygen or nitrous oxide
– Continuous operation with short-time
loading
WARNING! Read the safety
information and all instructions to
avoid serious injury.