YAS02CI (Option: L8)
!The external power supply must be in accordance
with EN 61010-1, Section 6: "Protection Against
Shock Current".
Please also refer to the specifications for
classification of electrically operated equipment
in EN 61010-1, Appendix H.
Safety Precautions
The power supply must be rated to safety extra
low voltage (SELV) or grounded (earthed) safety
extra low voltage (SELV-E).
The 24V industrial power grid module meets
following requirements:
EN 61000-4-6: Conducted Radio Frequency
0.15 MHz-80 MHz, 3 V, 80%AM (1kHz)
EN 61000-4-4: Burst
+/–2KV 5/50 tr/tn ns, 5 kHz
EN 61000-4-5: Surge Line to Line
+/–1KV 1.2/20 (8/20) tr/th us
EN 61000-4-5: Surge Line to Earth
+/–2KV 1.2/20 (8/20) tr/th us
The device is equipped for the following input
voltage ranges:
Power supply with direct current:
Input voltage range:
Power consumption:
Power supply with alternating current:
Input voltage range:
Frequency:
Power consumption:
Power supplied by car battery with Accessory
YCC02-CB02:
Input voltage range:
24 to 15.5 V +/–10%
12 W
17 to 13 V +/–10%
50/60 Hz
12 W
13 to 11 V