What to do if...
Problem
Battery Icon is flashing
Err 1
Err 2
No display
Measurements appear
too high or too low
Prior to measurement,
the positioning system
(A) on the display does
not work.
Reason
• Batteries are flat.
• Memory readings can be
called up, but measure-
ments are not possible.
Cuff is not wrapped on
the wrist well (too loose).
• Arm moved during
inflation measurement.
• Systolic pressure is
above 280mmHg
No information on the
display screen.
Blood pressure varies
constantly during the
day. Many factors such
as stress, time of day,
talking, coughing, laugh-
ing, moving etc. when
measuring will influence
the reading.
Positioning system (A) is
deactivated.
13
Solution
Insert new batteries.
Rewrap the cuff tighter
on the wrist.
Please do not move arm
or speak when you take a
measurement.
Check the battery instal-
lation for proper instal-
lation of the batteries'
polarities
Please read "Key rules for
accurate blood pressure
measurements" and
"Taking a measurement"
Refer to the chapter
«Select mode» to activate
the positioning system
(A).