1.1 Install the iChiropro app
A. Open the App Store ©, tap the corresponding icon on the iPad.
B. Tap Search to find the iChiropro application.
C. Install the iChiropro application designed by Bien-Air Dental.
Note 1
1.2 Install the iChiropro system
A. If necessary, use the Phillips screwdriver to change the
adapter for the iPad model being used (see chapter "11.5 iPad
adapter change" on page 50).
FIG. 1
B. Place the iChiropro on a flat surface capable of bearing its
weight.
⚠
CAUTION
It may be positioned on a table, on a trolley or any other surface
but in no circumstances on the floor.
FIG. 2
C. The fuse box may be opened with a screwdriver.
100 - 240 VAC = fuse T4.0AH 250 VAC REF 1307312-010.
To replace a fuse, see chapter"11.6 Fuses replacement" on page 51.
D. Connect the power cable (1) to the connector (2).
Note 2
⚠
CAUTION
The power plug is the device used for disconnection in case of
problems, it must be easily accessible at all times.
FIG. 3
E. Connect the pedal cable to the output provided on the rear panel,
guiding the connector and plug by means of the index pin on the
connector.
⚠
CAUTION
Do not lift the pedal holding the connection cable.
To disconnect the pedal cable pull the cable socket connector (1).
Do not pull the cable (2) without disconnecting the cable socket
before.
FIG. 4
F. Connect the MX-i micromotor cable to the motor output, guiding
the connector and plug by means of the index pin on the
connector.
FIG. 5
G. Align and attach the bracket to the housing provided on the rear
of the console and suspend the flask or bottle.
FIG. 6
H. Check the packaging integrity, as well as the expiry date of the
irrigation line on the label (1).
⚠
WARNING
The medical device must be used only with lines supplied by
Bien-Air Dental to ensure trouble-free operation. These lines are
sterile and for single use. Re-use may result in microbiological
contamination of the patient.
I. Remove the single-use sterile irrigation line (2) from its pouch.
FIG. 7
J. Connect the flexible hose of the irrigation line to the spray tube of
the handpiece or contra-angle.
FIG. 8
K. Install the peristaltic cassette (1) in the peristaltic pump (2).
Check that the cassette is clipped correctly.
FIG. 9
L. Close the pump lid (3). If there is resistance to closing, open the
lid again and check the correct positioning of the cassette. When
the lid is correctly closed, the user should hear a click sound.
⚠
CAUTION
Do not run the pump while the lid is open.
⚠
WARNING
Risk of pinching!
FIG. 10
M. Perforate the cap of the physiological liquid flask with the pointed
end of the irrigation line after removing the protective cap.
FIG. 11
N. Attach the irrigation line on the motor cable using the attachment
collars (1) REF 1303711-010.
1.3 Installation of the iPad on the iChiropro
O. Remove the single-use sterile protection sheet for iPad from its
pouch and stick it on the screen. Refer to the instructions on the
back of the pouch.
FIG. 12
P. Connect the iPad to the iChiropro by sliding it carefully along the
adapter.
1.4 On/off procedure
The device can be switched on and off in complete safety using
the main switch on the iPad and iChiropro.
NOTES
1
The iPad must be correctly connected to the Internet before
the App Store © is opened; refer to Apple's user guide for appro-
priate use of the iPad.
2
The equipment is powered by the mains power supply (100 -
240 Vac).
3