Manufacturer's declaration - electromagnetic immunity
VISION ELECT HDTV (model 240-030-960) monitor is intended for use in the electromagnetic
environment specified below. The customer or the user of VISION ELECT HDTV (model 240-030-960)
monitor should assure that it is used in such an environment.
Immunity Test
Electrostatic Discharge
(ESD)
IEC61000-4-2
Electrical fast transient/
burst
IEC61000-4-4
Surge
IEC61000-4-5
Power frequency
(50/60Hz) magnetic field
IEC 61000-4-8
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC61000-4-11
NOTE: Ut is the A.C. mains voltage prior to application of the test level.
IEC 60601
Test Level
6kV contact
8kV air
2kV for power
supply lines
1kV for input/output
lines
1kV differential
mode
2kV common mode
3.0 A/m
<5% Ut (>95% dip
in Ut) for 0.5 cycle
40% Ut (60% dip in
Ut) for 5 cycles
70% Ut (30% dip in
Ut) for 25 cycles
<5% Ut (>95% dip
in Ut) for 5 sec.
Compliance Level
6kV contact
8kV air
2kV for power
supply lines
1kV for input/output
lines
1kV differential
mode
2kV common mode
3.0 A/m
<5% Ut (>95% dip
in Ut) for 0.5 cycle
40% Ut (60% dip in
Ut) for 5 cycles
70% Ut (30% dip in
Ut) for 25 cycles
<5% Ut (>95% dip
in Ut) for 5 sec.
Electromagnetic
environment -
guidance
Floors should be
wood, concrete, or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least
30%.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of VISION
ELECT HDTV (model
240-030-960) image
intensifier requires
continued operation
during power mains
interruptions, it is
recommended that
VISION ELECT
HDTV (model 240-
030-960) image
intensifier be
powered from an
uninterruptible power
supply or a battery.