C
D
HEMIC AL
ISINFECTION
Testing by ResMed has shown that the mask components
can withstand up to 15 cycles of disinfection with Cidex†
Plus or Cidex PA when the manufacturer's instructions are
followed.
1
Disinfect the components using Cidex Plus or Cidex PA
and rinse. Follow the manufacturer's instructions.
2
Inspect the components for wear or deterioration.
Replace the affected part if damaged.
R
EASSEMBLING THE
1
Push the air vent into the mask frame so that the small
round bump on the vent is on the outside of the frame.
2
Fit the grooved edge of the nasal cushion to the mask frame.
Fit the top section first, then work your way around the
mask, pushing the cushion onto the frame, ensuring a good
seal has been achieved.
4
M
ASK
3
Replace the forehead pads by sliding the centre division of
the pads onto the tabs of the adjustable arms.
4
Refit the port plug on the forehead assembly. Connect the
forehead assembly to the mask frame and the short length
tubing to the forehead assembly.