Enclosure 2 – Assembling Smartact tacks on the device
Photo. 15
Open the sterile Smartact Pin blister pack (ar-
ticle code 4730).
Photo. 18
Open the Smartact Pin container.
Photo. 21
The Smartact system is ready to use.
Enclosure 3 – Assembling the compressed air connection kit
Photo. 22
Insert the ¼ female connector, the straight cylindrical plug con-
nector and then the 6 mm ¼ hose supplied with the kit.
Smartact
Photo. 16
Take the disposable tack insertion device from the
blister pack and screw it up on the curved body us-
ing the special torque wrench supplied with the kit.
Photo. 19
Insert the disposable tack insertion device in
the special housing and pick up the tack (pho-
tos 21).
Photo. 23
The assembled system can now be connected to the foot-
switch plug connector. Connect the system to the compressed
air supply of the practice; if necessary use a pressure adapter
to ensure an operating pressure between 2 and 6 bar.
Smartact
Photo. 17
Make sure you engage the thread correctly and
that the tack insertion device is screwed up fully.
Photo. 20
Tack inserted in the disposable tack insertion
device.
9