Section 8: Specifications

Guidance and Manufacturer's Declaration - Emissions
The 3M™ Bair Hugger™ temperature monitoring system is intended for use in the electromagnetic environment
specified below. The customer or user of the temperature monitoring system should ensure that it is used in
such an environment.
Emissions Test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonics
IEC 61000-3-2
Flicker
IEC 61000-3-3
Guidance and Manufacturer's Declaration – Emissions
The 3M™ Bair Hugger™ temperature monitoring system is intended for use in the electromagnetic environment
specified below. The customer or user of the temperature monitoring system should ensure that it is used in
such an environment.
Immunity
IEC 60601 Test
Test
Level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
Guidance and Manufacturer's Declaration – Immunity
The 3M™ Bair Hugger™ temperature monitoring system is intended for use in the electromagnetic environment
specified below. The customer or user of the temperature monitoring system should ensure that it is used in
such an environment.
16
Compliance
Group 1
Class B
Class A
Complies
Compliance
Level
3Vrms
3V/m
Electromagnetic Environment – Guidance
The temperature monitoring system uses RF energy only
for its internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby
electronic equipment.
The temperature monitoring system is suitable for use in
all establishments, including domestic, and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Electromagnetic Environment – Guidance
Portable and mobile communications equipment
should be separated from the temperature
monitoring system by no less than the distances
calculated/listed below:
D=1.2
D= 1.2 (80 to 800 MHz)
D= 2.3 (800 MHz to 2.5 GHz)
Where P is the max power in watts and D is the
recommended separation distance in meters.
Field strengths from fixed transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance levels
(V1 and E1).
Interference may occur in the vicinity of equipment
containing a transmitter.