11 General Setup (PATH: Menu/Quick Setup) continued
5.0
0.16
Aout1 min= 0.000 ppb
Aout1 max= 100.0 ppb
5.0
0.16
a Set Point Yes
SP1 Type= High
5.0
0.16
SP1 High = 0.000 ppb
– USP (% safety margin below U.S. Pharmacopoeia limits)
– EP PW (% safety margin below European Pharmacopeia limits for Purified Water)
– EPWFI (%safety margin below European Pharmacopeia limits for Water for Injection)
– JP Cond (% safety margin below Japanese Pharmacopoeia limits)
5.0
0.16
SP1 use Relay #1
UV Lamp
Once flow has been established, the 4000TOCe is ready to begin taking TOC meas-
urements. If sensor activation is desired at this time, set the UV Lamp to On. Allow
approximately one minute for readings to begin. Additional time (approximately 4-24
hours) may be necessary for the readings to stabilize as the sample line rinses clean,
and the TOC sensor reaches thermal stability.
Auto Start
The 4000TOCe sensor can be configured such that the UV lamp will automatically turn on
after recovering from a power loss or fault condition. If automatic recovery is desired, set
the Auto Start option to 'Yes'.
It is recommended that Auto Start be set to 'Yes'. The default setting is 'No'.
Analog outputs
ppbC
By selecting Yes the linear 4–20 mA analog output Aout1
µScm
will be set up for measurement a when [ENTER] is pressed.
ì
Selecting No means that no analog output is set up.
Aout1 min, Aout1 max are the minimum and maximum measu-
rement values for the 4 and 20 mA values respectively.
Set Points
ppbC
After configuring the Analog Output a setpoint can be configured
µScm
for that output. If No is selected and [ENTER] is pressed then the
quick setup is done and the menus are exited without setting
ì
up any setpoint.
Selecting Yes and pressing [ENTER] means a setpoint can be
ppbC
configured for measurement a.
µScm
Following setpoint Types can be selected:
ì
– High (High value has to be set)
– Low (Low value has to be set)
– Between (High and Low value has to be set)
– Outside (High and Low value has to be set)
After setting the set point value(s) a Relay (none, 1, 2, 3, 4)
ppbC
can be configured for that setpoint. The Relay delay is set to
µScm
10 seconds and the Hysteresis is set to 5%.