Disposal; Specifications - Mettler Toledo 4000TOCe Guia De Inicio Rapido

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 59

12 Disposal

When the sensor is finally removed from service, observe all local environmental
regulations for proper disposal.

13 Specifications

TOC Sensor
Measurement Range
Accuracy
Repeatability
Resolution
Analysis Time
Initial response time
Limit of Detection
Conductivity Sensor
Conductivity Accuracy
Cell Constant Accuracy
Temperature Sensor
Temperature Accuracy
Sample Water Requirements
Temperature
Particle Size
Minimum Water Quality
Flow Rate
Pressure
General Specifications
Case Dimensions
Weight
Enclosure material
Enclosure rating
Ambient Temperature/ Humidity rating
Power requirements
Local Indicators
Ratings/Approvals
12
0.05 - 1000 ppbC (µgC/L)
± 0.1 ppbC for TOC < 2.0 ppb (for water quality > 15 M�-cm [0.067 µS/cm])
± 0.2 ppbC for TOC > 2.0 ppb and < 10.0 ppb (for water quality > 15 M�-cm
[0.067 µS/cm])
± 5% of measurement for TOC > 10.0 ppb (for water quality 0.5 to 18.2 M�-cm
[2.0 to 0.055 µS/cm])
± 0.05 ppbC < 5 ppb, ± 1.0% > 5 ppb
0.001 ppbC (µgC/L)
Continuous
< 60 seconds
0.025 ppbC
± 2%,0.02-20 µS/cm; Constant Sensor*
± 2%
Pt1000 RTD, Class A
± 0.25°C
0 to 100 °C **
<100 micron
> 0.5 M�–cm (< 2 µS/cm), pH < 7.5 ***
20 mL/min
4 to 200 psig (0.3 bar(g) to 13.6 bar(g)) at sample inlet connection ****
11" [280mm] W x 7.4" [188mm] H x 5.25" [133mm] D
5.0 lb. (2.3 kg)
Polycarbonate plastic, flame retardant, UV and chemical resistant
UL # E75645, Vol.1, Set 2, CSA #LR 49336
NEMA 4X, IP65 Industrial environment
5 to 50°C / 5 to 80% Humidity, non-condensing
100 - 130VAC or 200 - 240VAC, 50/60 Hz, 25W Maximum
Four LED lights for Fault, Error, Sensor Status and UV Lamp ON
CE Compliant, UL and cUL (CSA Standards) listed, Conductivity and temperature
sensors traceable to NIST, ASTM D1125 and D5391. Meets ASTM D5173
Standard Test Method for On-Line Monitoring of Carbon Compounds in Water by
UV Light Oxidation

Publicidad