When the AUTO OFF feature is enabled, the stethoscope will enter standby mode after a period of time determined by the
POWER setting if there is no contact detected on the diaphragm AND the buttons are not activated. When the AUTO OFF
feature is disabled, the stethoscope will enter standby mode after a period of time determined by the POWER setting from
the last button activation. In addition, when the AUTO OFF feature is disabled, there will be an audible alert 10 seconds prior
to entering standby mode. After entering standby mode, the Model 3100 will be powered on when contact is detected on the
diaphragm or by button activation. Otherwise, the Model 3100 will completely power off after the period of time determined
by the POWER setting (see below).
With AUTO OFF feature ENABLED:
SETUP
Options
Detail
POWER
HIGH
Remains on full power for 30 seconds before entering standby
Remains in standby for 5 hours before powering off
Backlight remains lit for 5 seconds after pressing the power button
MEDIUM
Remains on full power for 20 seconds before entering standby
Remains in standby for 2 hours before powering off
Backlight remains lit for 5 seconds after pressing the power button
LOW
Remains on full power for 10 seconds before entering standby
Remains in standby for 30 minutes before powering off
Backlight remains lit for 3 seconds after pressing power button
With AUTO OFF feature DISABLED:
SETUP
Options
Detail
POWER
HIGH
Remains on full power for 7 minutes before entering standby
Remains in standby for 5 hours before powering off
Backlight remains lit for 5 seconds after pressing power button
MEDIUM
Remains on full power for 5 minutes before entering standby
Remains in standby for 2 hours before powering off
Backlight remains lit 5 seconds after pressing power button
LOW
Remains on full power for 3 minutes before entering standby
Remains in standby for 30 minutes before powering off
Backlight remains lit 3 seconds after pressing power button
12
To obtain model and software version information: Select the "ID" option in the Setup menu. Use the (-) and (+) buttons
to scroll through the list of reference numbers. Press (M) to return to the operating mode of the stethoscope.
To leave the Setup menu: Press the Filter button or press the (M) button to select the EXIT option. This will return the
stethoscope to its normal operating state.
SETUP
Id
exit
power
10. Other Operating Considerations
Operating range is -22º to 104ºF (-30º to 40ºC), 15 to 93% relative humidity.
Storage and transport range is -40º to 131ºF (-40º to 55ºC), 15 to 93% relative humidity.
To extend the life of your stethoscope, avoid extreme heat, cold, solvents and oils. Remove the battery whenever the
stethoscope will not be used for several months.
If you plan to use the stethoscope below 0ºF (-18ºC) you should use a lithium battery to insure proper function.
Failure to follow care and maintenance recommendations could result in damage to the internal components of the Littmann
Electronic Stethoscope. Internal damage could cause malfunction of the product, ranging from a slight decrease in auditory
response to complete failure of the product.
If you experience any problems with the electronic stethoscope, do not attempt to repair it yourself. Please notify our 3M
Health Care Service Center for directions on shipping and receiving.
13