TROUBLESHOOTING
If any abnormality arises during use, please check the following points:
SYMPTOMS
Unit does not turn on
Batteries have
when the
run down.
START/STOP
button is pushed.
Battery polarities have
been positioned
incorrectly.
EE measurement
Cuff has been placed
error symbol shown
incorrectly.
on display or the
blood pressure value
Did you talk or move
is displayed exces-
during
sively low
measurement?
(or high).
Shaking of the wrist
with the cuff on.
E1 error symbol
Air circuit
shown on display
abnormality.
E2 error symbol
Inflation pressure
shown on display
exceeding 300 mmHg.
E3 error symbol
Data error.
shown on display
"
--" appears on
Recalibration alert.
display
Note: If the unit still does not work, contact HoMedics Consumer
Relations. Under no circumstance should you disassemble or
attempt to repair the unit by yourself. Contact information for
HoMedics Consumer Relations Department can be found on
the warranty page (See page 35).
POSSIBLE
CORRECTION
CAUSES
Replace them
with two new AAA
alkaline batteries.
Re-insert the
batteries in the
correct positions.
Wrap the cuff properly
so that it is positioned
correctly.
Measure again.
Keep wrist steady
during measurement.
Measure again.
Switch the unit off,
then measure again.
Remove and reload the
batteries.
Refer to page 28 for
more information.
33
SPECIFICATIONS
P P o o w w e e r r S S o o u u r r c c e e : :
Two 1.5V DC (AAA) Alkaline batteries
M M e e a a s s u u r r e e m m e e n n t t
Oscillometric
M M e e t t h h o o d d : :
Pressure: 40~250 millimeters Mercury
M M e e a a s s u u r r e e m m e e n n t t
R R a a n n g g e e : :
Pulse: 40~199 beats/minute
Pressure: ±3 mmHg
A A c c c c u u r r a a c c y y : :
Pulse: ±5% of reading
P P r r e e s s s s u u r r e e S S e e n n s s o o r r : :
Semi-conductor
I I n n f f l l a a t t i i o o n n : :
Pump driven
D D e e f f l l a a t t i i o o n n : :
Automatic Pressure release valve
M M e e m m o o r r y y C C a a p p a a c c i i t t y y : : 30 memories for each User (60 total)
A A u u t t o o - - s s h h u u t t - - o o f f f f : :
1 minute after last button operation
O O p p e e r r a a t t i i o o n n
Temperature: 50°F~104°F (10°C~40°C)
E E n n v v i i r r o o n n m m e e n n t t : :
Humidity: 40%~85% RH max.
S S t t o o r r a a g g e e
Temperature: 14°F~140°F (-10°C~60°C)
E E n n v v i i r r o o n n m m e e n n t t : :
Humidity: 10%~90% RH max.
W W e e i i g g h h t t : :
0.25 lbs (without batteries)
W W r r i i s s t t
5.3"- 8.6" (13.5 - 22cm)
C C i i r r c c u u m m f f e e r r e e n n c c e e : :
3.25"(L) x 2.75"(W) x 2.60"(H)
D D i i m m e e n n s s i i o o n n s s : :
82mm(L) x 69mm(W) x 66mm(H)
Type BF
Device and cuff are designed to
provide special protection against
electrical shocks.
Note: These specifications are subject to change
without notice.
(mmHg)
34