M400 Transmitter: New Functions
EN M400 Transmitter Series: New Functions
1
Concentration Table
If during the channel setup (see chapter 8.2.1 "Channel setup") the parameter
conductivity has been chosen or an four-electrode conductivity sensor based
on ISM technology is connected to the transmitter, a concentration table can
be defined.
To specify customers-specific solutions, up to 9 concentration values can be
edited in a matrix together with up to 9 temperatures. To do so the desired
values are edited under the concentration table menu. Furthermore the
c onductivity values for the according temperature and concentration values
are edited.
For doing the settings the menu "Concentration Table", that will be displayed,
has to be chosen. (see chapter 8.2.3 "Parameter related settings").
NOTE: Refer to section 8.2.1 "Channel Setup" to choose the unit used
h
in the display.
© Mettler-Toledo AG, CH-8606 Greifensee
Define the desired unit.
Press [ENTER].
Enter the amount of desired temperature
points (Temp Point) and Concentration
Points.
Press [ENTER].
3