Descargar Imprimir esta página

Technical Specifications - Miniland thermoadvanced pharma Manual De Instrucciones

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 1
ENGLISH
6.1. CALIBRATION
The thermometer is calibrated during its manufacturing process. If the
thermometer is used following the instructions provided in the manual,
periodic re-calibration is not necessary. If at any time you have doubts
about the accuracy of the measurement, immediately contact the
establishment where you purchased the thermometer.
Do not attempt to modify or disassemble the thermometer.

7. TECHNICAL SPECIFICATIONS

7.1. TECHNICAL SPECIFICATIONS OF THE THERMOMETER
Product
Model
Power supply
Operation mode
Body reference site
Temperature detection part
Body temperature
measurement range
Laboratory accuracy of body
measurement
Object temperature
measurement range
Accuracy of object
measurement
Room temperature
measurement range
Accuracy of the room
measurement
Measuring time
Measuring interval
Non-contact forehead thermometer
TH1009N-C (thermoadvanced pharma
89379)
3 V DC (2 AAA batteries)
Adjusted mode
Heart
IR sensor
32.0°C ~ 42.9°C (89.6°F ~ 109.3°F)
± 0.2 °C (or ± 0.4 °F) from 35.5 °C - 42.0
°C (or 95.9 °F - 107.6 °F)
± 0.3 °C (± 0.5 °F) for other ranges
-22.0°C ~ 80ºC (-7.6°F ~ 176°F)
± 2.0 °C (± 4.0 °F)
5.0°C ~ 59.9°C (41.0°F ~ 139.8°F)
± 1 °C (± 2°F)
1 second
1 second
34

Publicidad

loading