PLEASE READ BEFORE USE
Failure to follow these instructions could render the device unusable and void any warranty.
Device Description
SUNOPTIC TECHNOLOGIES
coupled to a medical grade fiberoptic light source. SUNOPTIC TECHNOLOGIES
cables can be used with quartz halogen, metal halide, or xenon light sources. They are
compatible with virtually all endoscopes, medical instruments, and microscopes.
Intended Use
This device is designed to illuminate a surgical site by relaying light from a fiberoptic light source
onto the desired site. SUNOPTIC TECHNOLOGIES
transmittance, peak efficiency.
NOTE: The style of cable suitable for a given lightsource can be determined from the table
below.
Cuda / Sunoptic
250W Halogen
Other 250W Halogen
Focused Ceramic
Xenon
Other Xenon ≤ 300W
Xenon >300W
Metal Halide ≤ 75W
For use with Xenon lightsources, be sure lightsource is filtering at least 90% of infrared
•
radiation (IR) to prevent high heat.
Contraindications
This device has no contraindications of which we have knowledge.
Warnings and Precautions
SUNOPTIC TECHNOLOGIES
•
before use. See instructions for cleaning and sterilization.
Always inspect cables for any evidence of damage prior to use. Pay particular attention to
•
optical surfaces looking for scratches or dings.
Use caution to treat SUNOPTIC TECHNOLOGIES
•
device.
The user of this product should be thoroughly familiar and trained in use and care of the
•
product.
Match the cable size to the endoscope. Using a larger cable (fiber bundle size) than the
•
endoscope requires will result in overheating the endoscope's proximal end.
LIT021 IN'TECH MEDICAL
(English)
®
fiberoptic cables are designed to deliver maximum light when
Standard Cable
Clad Rod Cable
X
X
X
®
light cables are provided non-sterile and must be sterilized
®
cables are medical grade, high
Lensed Cable
X
X
X
®
cables as you would any fine optical
Date of Revision:5/23/13
®