Cleaning, disinfection, sterilization

5.1

Safety instructions

May be cleaned in an ultrasonic bath.
Never use sharp objects for cleaning.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 2 bar.
Maximum water pressure during rinsing: 2 bar.
During machine cleaning and disinfection, do not exceed 95 °C.
It is essential to follow the cleaning agent and disinfectant man-
ufacturers' instructions.
During sterilization, do not exceed 138 °C.
Do not sterilize in hot air.
5.2

Reconditioning limitation

Frequent reconditioning can damage this product. Observe the
safety instructions for checking the product. With visible damage
or functional impairment, the product may no longer be used.
5.3

Recommended equipment / substances

The service life of the product, cleaning action and sterility are
validated for the following cleaning and preparation procedures
if the recommended or equivalent substances are used in accor-
dance with the manufacturer's instructions.
• Manual cleaning/disinfection with the detergent
Cidezyme ® and disinfectant Cidex ® OPA (Johnson & John-
son Medical Limited, Gargrave, Skipton).
• Machine cleaning and disinfection in a G 7836 CD disinfector
(Miele & Cie. KG, Gütersloh, Germany) with the cleaning