4
How to use
4.1
Only use cleaned, disinfected and sterilized
products
Clean, disinfect and sterilize the product before using it for the
first time and before every subsequent use.
4.2
Check the product
Before every use check the insulation of the connecting cables,
the activation functions of the product and whether the product
has any physical damage.
If damaged, do not use this product!
4.3
Assembling the product
Insert the silver ball electrode (art. no. 20191-079 or 21191-
052) into the electrode handle Nervtest (art. no. 20190-148).
4.4
Connect up the product
Connection to an ICC 350 ZMK:
Insert the handle into the Neurotest receptacle on the Erbe ICC
350 ZMK.
Connect the remote control to the matching socket on the back
of the Erbe ICC 350 ZMK.
Using the remote control, select the required level (1 to 5) for
current amplitude before you activate the Neurotest device.
Connection to an NT 2:
To connect the handle, please read the respective chapter in the
NT 2 user manual.