Fault detected
Device does not operate
(LED indicator off)
No spray
Power fault/
low vibrations
No ultrasonic vibration
Leak between the
handpiece cord and the
connector on the device
Leak of liquid under the
device
Leak of liquid at the tank
connector level.
Possible causes
Power cord not plugged in properly.
Power power switch in position O.
Power receptacle fuse(s) blown.
Internal fuse blown.
Irrigation solution tank empty.
Faulty pump.
Irrigation deactivated
Blocked irrigation circuit
Blocked tip or file.
Wrong choice of tip.
Tip worn or distorted.
Incorrect power adjustment.
Incorrect use: incorrect angle of attack or
inadequate pressure on the tooth..
Liquid or moisture present between
handpiece and cord.
Incorrect tip tightening.
Faulty connector contact.
Handpiece cord wire cut.
Worn handpiece 1.15x1 seal.
Tube rupture.
Worn seals.
Solutions
- Check the power socket.
- Return to SATELEC after-sales.
- Place the power switch in position I.
Replace the power receptacle fuses.
Return to SATELEC after-sales.
Fill the irrigation solution tank.
Replace the pump.
Return to SATELEC after-sales
Return to SATELEC after-sales.
Unblock the tip or file.
Check the tip.
Replace the tip.
Refer to the TipBook for the necessary
information.
Refer to the TipBook for the necessary
information.
Dry the electrical contacts thoroughly.
Tighten the tip with the wrench.
Clean the connector contacts.
Return to SATELEC after-sales to replace
the cord.
Replace the seal (kit F12304).
Return to SATELEC after-sales.
- Replace the two seals (E15121).
- Return to SATELEC after-sales.
15