Descargar Imprimir esta página

A.5.1 Bit Speicherbereich - sylvac D302 Manual Del Usuario

Conversión del módulo de sensores

Publicidad

Access to a variable may be:
Read and write (R/W)
Read only (RO)
Write only (WO)
R/W* indicates that a variable can be read but always returns 0.
RO∆ indicates that the value can only be read once (reading causes the variable to be reset
to zero)
Access to a probe position variable while the probe is not connected returns the code -1 (not a
number, NAN in the usual IEEE754 float format).
A.5.1 Address of the Modbus bit variables
Function
Freezes the measurement
Min., max. and delta reset
to zero
Preset recall
Preset on Min value
Preset on Max value
Probe activation
Measurement in inches
Absolute measurement
Negative measurement
direction
Activated tolerances
Printing of activated tole-
rances
Active diameter mode
Correction point per point
activated
Stored correction point per
point
Reset of correction point
per point
Freezes the measurement
of all channels.
RS-485 bit length
RS-485 stop bits
RS-485 parity odd/even
RS-485 parity on/off
Global foot pedal input
Global foot pedal output
A
Probe synchronization on/off
Temperature of module
Address
Nb of bits
(decimal)
0000
1
0001
1
0002
1
0003
1
0004
1
0064
1
0065
1
0066
1
0067
1
0068
1
0069
1
0070
1
0074
1
0128
1
0192
1
8000
1
8128
1
8129
1
8130
1
8131
1
8133
1
8134
1
8135
1
8576
1
62
Access type
Comment
R/W
0 to release
R/W*
R/W*
R/W*
R/W*
R/W
0 to disable
R/W
0 for measurement in mm
R/W
0 for relative measurement
R/W
0 for positive direction
R/W
0 for disabled tolerances
R/W
0 for disabled printing
R/W
0 for inactive diameter mode
R/W
1 to enable, 0 to disable correction
0 : no stored correction
RO
1 : stored correction
R/W*
1 to reset the correction
R/W
0 to release
R/W
0 : 7 bits data - 1 : 8 bits data
R/W
0 : 1 stop bit - 1 : 2 stop bits
R/W
0 : Odd / 1 : Even
R/W
0 : without parity / 1 : with parity
R/W
0 : inactive input / 1 : active input
R/W
0 : inactive output / 1 : active output
R/W
0 : without synchro / 1 : with synchro
RO
Bit set = temperature stable